CHEMISTRY
AND THE AQUARIUM by RANDY HOLMES-FARLEY, Ph.D.
Sponsored
in part by:
Purity
of Calcium Chloride
Proud
sponsor of this column
Calcium chloride
is a compound of significant interest to many marine aquarists.
Besides its potential use in formulating artificial salt mixes,
it is also useful in directly supplementing
calcium to aquaria. It is sold by many manufacturers for
aquarists, and is also sold by companies outside of the reef
aquarium hobby for a wide variety of uses, ranging from melting
ice on sidewalks to formulating pharmaceuticals.
Because of these
varied uses and manufacturers, aquarists wanting to use calcium
chloride are presented with many options concerning its purchase.
Aquarists often ask whether it is necessary or desirable to
purchase the highest quality and most expensive grades of
calcium chloride. Until recently, I typically answered that
I did not know, but that I would avoid the lowest (least expensive)
grades of calcium chloride (often called Technical, Practical,
or with no grade described at all).
In a recent round
of studies, I purchased several commercial brands of calcium
chloride, and tested them for the impurities that I thought
likely to be a potential concern. The brands tested were Warner Marine
(Concentrated Calcium Supplement), ESV
(Calcium Chloride), Kent
(Turbo Calcium and Liquid Calcium), and Dow (Dowflake
77-80% Calcium Chloride; purchased from Home Depot). The sections
below detail the impurities that were found in each of these
brands. It also reviews the results in terms of how important
these impurity levels might be, and compares them to each
other and to the impurities that would be delivered using
commercial calcium carbonate in a calcium carbonate/carbon
dioxide reactor to deliver the same amount of calcium.
As a follow up
article, next month I will show how to make your own two part
additive using calcium chloride, baking or washing soda, and
Epsom salts.
ICP Testing
Methods
Feel free to skip
to the last line (in bold) of this method section if testing
details do not interest you.
There has recently
been a significant amount of discussion in the hobby concerning
analytical methods. Richard Harker has recently authored two
articles that demonstrate some concerns with a particular
method involving ICP (Inductively Coupled Plasma;
Article 1,
Article 2). In this technique, the liquid sample is injected
into an incredibly hot plasma, completely breaking the sample
down into individual elemental ions. There are a variety of
ways to then detect these ions, including mass spectroscopy
(called ICP-MS) and from the intensity of different wavelengths
of light emitted from the different ions (called ICP-OES,
with OES standing for Optical Emission Spectroscopy).
I have chosen
to use ICP-OES, largely because it is the instrument that
I have available (A Varian Vista-MPX). It is also the method
that Richard Harker has claimed to be problematic when using
a particular protocol (ICP Scan, EPA method 200.7). I share
some of his concerns about the use of ICP, particularly for
ions that are near their detection limits. In fact, I believe
that some analyses reported in the aquarium literature using
ICP-OES may simply reflect instrumental and sample noise that
was added up by the software used with the instrument to report
an artificially high value. I have discussed this concern
in the past, for example, for aluminum.
In all ICP-OES
data that I have presented in prior articles, as well as all
data in this article, such issues are not a consideration
for the reader. I have carefully looked at the emission data
for each ion and for each sample, with my own eyes, to confirm
whether the data claimed by the software is a real emission,
or just noise. For all signals near the noise level, I spiked
small but known amounts of commercial ICP standards into these
calcium chloride solutions to see exactly how high of a concentration
would have to be present in these solutions to be able to
see real signals. These detection limits are often different
than the detection limits stated by the instrument manufacturer
(since those stated values do not take into account the nature
of other interfering ions in your test sample, such as a huge
background of calcium and chloride), and are shown later in
the article. I have also tested the highly purified water
used to dissolve the solid samples in this study. This type
of protocol takes much more effort than running the sort of
ICP scan that Harker has discussed. Nevertheless, it is important
when trying to understand the limits of the data obtained.
This process
is shown in Figure 1. Figure 1 is a plot of the emission intensity
in one of the wavelength regions expected to have emissions
from cadmium. The bottom spectrum in Figure 1 is part of the
emission from the Dow sample. There is no apparent peak at
228.802 nm where one expects one of the emission peaks for
cadmium. However, when 0.1 ppm cadmium is added to this sample
(top spectrum in Figure 1), the cadmium emission is clearly
seen. Adding 0.02 ppm Cd instead of 0.1 ppm results in a smaller,
but still clearly defined peak above the background noise.
From those results, I conclude that the Dow sample has less
than 0.02 ppm cadmium.
Figure 1. Emission spectrum in the region of one of the
emissions from cadmium. The bottom curve represents the
emission from the Dow sample. The upper curve represents
this same sample spiked with 0.1 ppm cadmium. The H-shaped
symbol in the middle represents the region actually integrated
to collect data (between the vertical lines).
Figure 2 shows
the cadmium emission from the Kent Turbo Calcium sample. Even
without any extra cadmium added, it has a clearly defined
cadmium peak at 228.802 nm. When that peak area is compared
to cadmium standards, I conclude that it has about 0.07 ppm
cadmium. This process was repeated for 2-3 emission peaks
for each ion and for each sample. The tabulated results represent
an average of these values for each ion. Table 1 shows the
limits of detection that I determined using the above process
for many of the ions tested in this article (for most of the
other ions, the limit of detection was far below the values
found).
Figure 2. Emission spectrum in the region of one of the
emissions from cadmium for the Kent Turbo Calcium sample.
The H-shaped symbol in the middle represents the region
actually integrated to collect data (between the vertical
lines).
In short, and
summarized for folks that do not have much interest in instrumental
methods:
The data in
this article do not suffer from any of the complications previously
discussed by Harker and others relating to ICP, and can be
considered reliable indications of what was present in the
samples tested.
Table
4. Experimental limit of detection in calcium chloride
solution by ICP.
Also
shown is the stated machine limit of Quantitation under
ideal conditions.
Element
Published
Machine Minimum
Limit of Quantitation (ppm)
Estimated
Experimental Limit of
Detection in CaCl2 Solution (ppm)
Arsenic
0.12
0.5
Cadmium
0.015
0.02
Cobalt
0.05
0.1
Copper
0.02
0.02
Chromium
0.04
<
0.02
Lead
0.14
0.5
Manganese
0.003
<
0.02
Molybdenum
0.04
0.1
Nickel
0.06
0.1
Selenium
0.37
0.5
Strontium
0.0002
<
0.02
Zinc
0.009
<
0.02
Sample Preparation
The samples that
were liquid (Kent Liquid Calcium and Warner Concentrated
Calcium Supplement) were tested as is. Samples that
were solid (Dow Flake 77-80% Calcium Chloride, ESV Calcium
Chloride and Kent Turbo Calcium) were dissolved in highly
purified laboratory deionized water (that was itself tested
for impurities) to approximately the same calcium levels found
in the liquid products (about 100,000 ppm calcium, or 10%
calcium by weight).
Proud
sponsor of this column
It should be noted
that the Dow sample is calcium chloride dihydrate (CaCl2
· 2 H2O), while the Kent and ESV solids are
anhydrous calcium chloride (CaCl2). That difference
is where the 77-80% comes in with the Dow material. It is
not that there is 20+% impurities there, but rather that it
contains more than 20% water (the exact formula CaCl2
· 2 H2O contains 24.5% water, and the Dow
material has presumably lost a bit of this water in manufacturing).
ICP Results
The results in
Table 1 show the results obtained by ICP for these calcium
chloride solutions. Colored entries indicate where there are
significant differences between the samples (which I usually
decided was at least a factor of 2 difference in impurity
concentration). It is not intended to indicate that there
is actually a problem with any of these values. That analysis
is more complicated, and follows in later sections of this
article. Some entries are shown as "< X", which should
be read as "less than X", and indicates that the element is
below the detection limit that I established for each ion.
It might be slightly less than X, or it might be far, far
less.
One interesting
attribute emerges from these data immediately, however. The
DOW, ESV and Warner products seem to have one impurity profile,
and the two Kent products have a different one. This difference
is particularly clear for potassium, with huge amounts in
the DOW, Warner, and ESV samples, and hardly any in the two
Kent products. Fortunately, there is already a very large
amount of potassium in seawater, and it is not especially
toxic, nor is it a concern to aquarists. There are, of course,
some additional differences that will be discussed below,
but this fundamental dichotomy may reflect fundamentally different
manufacturing processes to arrive at calcium chloride.
Table
1. Experimental concentrations in liquid solutions (ppm)
Element
Dow
Solid
Kent
Liquid
Kent
Solid
ESV
Solid
Warner
Liquid
Aluminum
0.18
0.15
0.11
0.19
0.16
Arsenic
<
0.5
<
0.5
<
0.5
<
0.5
<
0.5
Barium
0.42
0.57
0.48
0.39
0.52
Beryllium
0.0013
0.0018
0.0016
0.0017
0.0011
Boron
20
0.4
0.6
14
20
Cadmium
<
0.02
0.07
0.07
<
0.02
<
0.02
Cobalt
<
0.1
<
0.1
<
0.1
<
0.1
<
0.1
Copper
<
0.02
<
0.02
<
0.02
<
0.02
<
0.02
Chromium
0.04
0.06
0.10
0.05
<
0.02
Lead
<
0.5
<
0.5
<
0.5
<
0.5
<
0.5
Lithium
38
<
0.1
<
0.1
37
34
Magnesium
11
3
19
4
7
Manganese
0.01
0.007
0.02
0.008
0.005
Molybdenum
<
0.1
<
0.1
<
0.1
<
0.1
<
0.1
Nickel
<
0.1
<
0.1
<
0.1
<
0.1
<
0.1
Phosphorus
<
0.2
<
0.2
<
0.2
<
0.2
<
0.2
Potassium
3600
7
5
3500
3200
Selenium
<
0.5
<
0.5
<
0.5
<
0.5
<
0.5
Silicon
1.4
1.0
1.3
1.1
1.5
Sodium
1015
28
25
920
660
Strontium
500
17
19
550
300
Vanadium
<
0.01
<
0.01
<
0.01
<
0.01
<
0.01
Zinc
0.040
0.011
0.015
0.015
0.012
In order to assess
whether these impurities are a real issue, one must understand
how much will actually be added to an aquarium, and how much
of a concern the final concentration in the aquarium is to
organisms in the aquarium. In order to make such an assessment,
Table 2 shows the concentration that will be added in total
over a year to an aquarium using these products. For this
analysis, the amount delivered was set at the equivalent of
16.2 ppm calcium per day, which is equivalent of adding 2%
of the tank volume in saturated limewater every day. This
level would be considered to be medium to high demand for
typical reef aquaria.
Table 2 also shows
data on the delivery of impurities using Koralith (commercial
calcium carbonate media) in a CaCO3/CO2
reactor, based on data provided in an
earlier article by Craig Bingman.
Greg Hiller has also analyzed other media, and I converted
data from both of these articles into delivery per year values
in
another article.
To assist in interpreting
some of these values, Table 3 shows the percentage increase
over the natural seawater levels that accrue over a year.
Table 3 only covers some of the major ions. I believe it would
be misleading to show such data for many of the trace ions
as there is little reason to believe that the concentrations
of such ions in aquarium water matche natural seawater, even
in the absence of any calcium chloride additions. So, for
example, it might be very misleading to say that copper additions
were ten fold higher than natural seawater levels, if the
aquarium water might already have 100 times those levels.
For each instance
where a clear distinction can be made, I have highlighted
in red those values in Table 2 that are appreciably higher
than the others for the same ion. Interestingly, Koralith
has more of these highlights than do the calcium chloride
samples. That comparison may not be complete (the calcium
chloride samples do not provide alkalinity, for example, so
other supplements must also be used), but it does suggest
that there are many ions for which calcium chloride is not
a problematic source.
Keying on those
red highlights for the calcium chloride samples, some can
clearly be discounted as being relatively unimportant, but
others cannot. Strontium,
for example, is not being delivered in adequate quantity by
any of these methods to match that likely to be removed by
calcification, so it is unlikely to rise to problematic levels
from this source. Boron
too is not being significantly raised over natural levels.
Table
2. Elements added in 1 year (ppm in the water).
Element
Dow
Solid
Kent
Liquid
Kent
Solid
ESV
Solid
Warner
Liquid
Koralith
CaCO3
Aluminum
0.011
0.009
0.006
0.011
0.009
1.0
Arsenic
<
0.03
<
0.03
<
0.03
<
0.03
<
0.03
0.05
Barium
0.03
0.03
0.03
0.02
0.03
0.000
Beryllium
0.00008
0.0001
0.0009
0.0001
0.00006
Not
tested
Boron
1.0
0.02
0.04
0.8
1.0
<
0.006
Cadmium
<
0.001
0.004
0.004
<
0.001
<
0.001
<
0.005
Cobalt
<
0.006
<
0.006
<
0.006
<
0.006
<
0.006
0.01
Copper
<
0.001
<
0.001
<
0.001
<
0.001
<
0.001
0.1
Chromium
0.002
0.004
0.006
0.003
<
0.001
<
0.007
Lead
<
0.03
<
0.03
<
0.03
<
0.03
<
0.03
0.1
Lithium
2.2
<
0.006
<
0.006
2.2
2.0
<
0.001
Magnesium
0.6
0.2
1.0
0.2
0.4
14.
Manganese
0.001
0.000
0.001
0.000
0.000
1.0
Molybdenum
<
0.006
<
0.006
<
0.006
<
0.006
<
0.006
<
0.006
Nickel
<
0.006
<
0.006
<
0.006
<
0.006
<
0.006
0.2
Phosphorus
<
0.01
<
0.01
<
0.01
<
0.01
<
0.01
<
0.08
Potassium
210
0.4
0.3
210
190
<0.06
Selenium
<
0.03
<
0.03
<
0.03
<
0.03
<
0.03
<
0.08
Silicon
0.08
0.06
0.08
0.07
0.09
1.1
Sodium
60
1.6
1.5
54
39
0.1
Strontium
29
1.0
1.1
32
18
0.8
Vanadium
<
0.001
<
0.001
<
0.001
<
0.001
<
0.001
0.02
Zinc
0.002
0.001
0.001
0.001
0.001
0.3
Table
3. Elements added in 1 year as a percentage of that
present in natural seawater.
Element
Dow
Solid
Kent
Liquid
Kent
Solid
ESV
Solid
Warner
Liquid
Koralith
CaCO3
Boron
23
0.4
0.9
18
23
<
0.1
Lithium
1200
<
3
<
3
1200
1100
<
0.5
Magnesium
0.05
0.02
0.08
0.02
0.03
1
Potassium
53
0.01
0.001
53
48
<
0.02
Sodium
0.6
0.02
0.01
0.5
0.4
0.001
Strontium
360
13
14
400
230
10
So what important
issues exist? That is hard to say with certainty. Four ions
seem worthy of further discussion: barium, cadmium, chromium
and lithium.
Barium, while toxic at elevated concentrations, would
not seem to pose a hazard at 0.02 - 0.03 ppm (which is only
2-3X over the natural level of 0.013 ppm). Most toxicology
studies do not show effects in marine organisms until well
above 1 ppm. Regulatory agencies (US and Canada) have not
set a limit for barium in seawater. So I conclude that the
barium levels in the calcium chloride samples are likely not
a problem.
Proud
sponsor of this column
Cadmium,
on the other hand, is known to be toxic at fairly low levels.
Canadian Water Quality Guidelines for the Protection of Aquatic
Life suggest a maximum of 0.12 ppb in seawater.
Toxicology tests on marine organisms show toxic effects
beginning in the 10-100 ppb range. Consequently, the delivery
of 4 ppb (for the two Kent products) over the course of a year
may be a concern. The other calcium chloride samples had significantly
less cadmium.
Chromium is another potential concern with some of these
samples.
Canadian Water Quality Guidelines for the Protection of Aquatic
Life suggest a maximum of 56 ppb in seawater for chromium
III compounds, and 1.5 ppb for chromium VI compounds. Chromium
compounds in an aquarium are likely to be chromium III as
the chromium VI compounds are highly oxidizing and will rapidly
react with organics and other inorganic ions to give chromium
III compounds again. Consequently, the delivery of 4-6 ppb
of chromium (for the two Kent products) over the course of
a year is not likely a significant concern.
Lithium does not seem to pose as much of a toxicity concern
as many other ions, but in three of the samples (Dow, Warner,
ESV) it is greatly elevated. Over a year, each of these would
add about 2 ppm lithium, or 12 times the natural level. Is
that too much? I am not sure. In a prior analysis of artificial
salt mixes, Craig Bingman found that two salt mixes started
out with greatly elevated lithium levels (90X over natural
levels for Coralife and 6X for Seachem, with the others ranging
from 1.5X - 3.1X). Typical aquaria surveyed by
Ron Shimek contained about 0.6 ppm of lithium (3X over
natural seawater) with a range from 0.015 ppm (0.08X) - 7
ppm (39 X).
It is well known
that excess lithium has significant adverse effects of the
development of sea urchin embryos,1,2 and many
studies have been carried out in this area. The amount of
lithium used in those studies, however, is typically around
500-3,000 ppm. It has also been shown that 345 ppm of lithium
will result in death of the isopod
limnoria.3
So while 2 ppm
lithium delivered by these samples is greatly increased over
the natural levels of 0.18 ppm, it is still small compared
to the hundreds of ppm required to show toxic effects. Given
that gap, and the fact that the lithium levels will likely
be attenuated by water changes, I conclude that the lithium
in these calcium chloride samples is not an excessive risk.
Nevertheless, that is something that each aquarist can decide
for themselves.
Ammonia Testing
There are several
ways that calcium chloride can be prepared on an industrial
scale. One of these (the Solvay process) involves ammonia.
Consequently, ammonia has the potential to be present as an
impurity in calcium chloride. For that reason, I tested each
of the calcium chloride samples for ammonia. I used two different
kits to test for ammonia: LaMotte and Red Sea. The results
of the Red Sea kit are shown in Table 4.
Samples spiked
with ammonia (from a standard containing 5.8 ppm ammonia as
ammonium hydroxide in water) did not show as much ammonia
as the test kit claimed, but it was clearly detectable in
the two spiked samples (Dow and Kent Turbo Calcium). These
spiked samples contained an extra 1.9 ppm of ammonia. One
showed up as 0.5 ppm ammonia, and the other showed as 0.5-1
ppm. Since all of the unspiked samples showed 0.5 ppm or less
of ammonia by the kit, I conclude that these samples have
less than 3 ppm of ammonia in them as tested (accounting for
dilution).
In short, none
of these samples showed enough ammonia to be concerned about,
even when adding enough to boost calcium by 200 ppm in one
day. Since these solutions were 100,000 ppm in calcium, adding
200 ppm calcium to an aquarium entails adding 1/500th
of the tank volume. At 3 ppm ammonia in the supplement, that
means that the tank will be boosted by 3/500 = 0.006 ppm of
ammonia, which I believe to be largely insignificant in a
reef aquarium.
Table
4. Ammonia in calcium chloride solutions.
Test
Fluid
DI
water (mL)
5.8
ppm ammonia standard
(mL)
Total
Ammonia shown
by Red Sea kit (ppm)
2
mL Dow
1
0
<0.25
2
mL Dow
0
1
0.5
2
mL Kent Solid Solution
1
0
<0.25
2
mL Kent Solid Solution
0
1
0.5
- 1.0
2
mL Kent Liquid
1
0
<
0.25
2
mL ESV
1
0
<
0.25
2
mL Warner
1
0
0.25
- 0.5
none
2
1
1
Conclusions
While I was unable
to test everything that I might like about these samples,
the tests reported here should be enough to allow folks to
make more informed decisions about calcium chloride. In particular,
the Dow Flake 77-80% Calcium Chloride does not look to be
appreciably worse than those sold by some of the aquarium
supply companies. Which brand is the best? I do not believe
that there is enough information presently available about
the toxicity of cadmium and lithium to really say whether
those that tend to have more of one or the other of these
is clearly best, so I will leave that to individual aquarists
to decide for themselves.
It should be noted
that these results only represent the samples that I actually
tested. Other batches from the same manufacturers may be different
(for better or worse). Likewise, these manufacturers may switch
to different sources for the calcium chloride at any time.
That consideration is one reason that I do not do more such
testing: it may not represent future product sold by the companies
involved. Given that concern, higher grades do have more quality
control checks, and so give more assurance that the product
will remain of acceptable quality in the future.
Proud
sponsor of this column
I do believe that
the Dow material will likely be acceptable in many applications,
and for those on a tight budget, it seems like a reasonable
choice. Toward that end, next month I will provide a recipe
for making an inexpensive two part additive system out of
the Dow calcium chloride (or any brand you prefer), baking
or washing soda, and Epsom salts (magnesium sulfate).
In the meantime,
Happy Reefing.
References
1. Effect of
lithium on ionic balance and polyphosphoinositide metabolism
during larval vegetalization of the sea urchin Paracentrotus
lividus. Ciapa, Brigitte; Maggio, Katia. Fac. Sci., Univ.
Nice, Nice, Fr. Developmental Biology (Orlando, FL, United
States) (1993), 159(1), 114-21.
2. Lithium
blocks cell cycle transitions in the first cell cycles of
sea urchin embryos, an effect rescued by myo-inositol.
Becchetti, Andrea; Whitaker, Michael. Dep. Phys. Sci., Univ.
Newcastle upon Tyne, Newcastle upon Tyne, UK. Development
(Cambridge, United Kingdom) (1997), 124(6), 1099-1107.
3. Lithium
as a substitute for sodium in Limnoria. Hinshaw, Lerner
B. Univ. of S. California, Los Angeles, Physiological Zoology
(1956), 29 81-5.